Setup
Change basic parameter
Change basic parameter is used to adapt the factory settings (name, unit,
display format) to your requirements. These details are used among others, for
calibration purposes.
Select the desired channel (e.g. analogue input A2) and then click Change ba-
sic parameter. Enter your adjustments
5
(e.g. Mbar instead of Volt).
To enter the conversion factor (e.g. Volts to Mbar) the following options are avail-
able:
Enter the amplification and the zero point
Using the calculation assistant
6
enter the values for points P1 and P2
7
.
The program will then calculate the amplification and the zero point.
Activate Apply user information by checking the checkbox
4
and then click
OK. Transfer the changes to the logger by clicking Write user settings.
Note: Should you wish to use a different display unit to that used for calibration,
then please use the function Change display parameter.
3
4
6
6
Max. 4 characters
Changes have been made to this channel.
7
5
Resetting channel properties to the factory settings:
Uncheck checkbox
4
(Apply user information) and transfer the changes to the
logger by clicking Write user settings.