8.3 Filter Maintenance
Depending on the application and environmental conditions, the filter must be checked from time to time and
cleaned or replaced when soiled.
1. Unscrew the filter from the probe housing and carefully pull the filter off the probe in a straight line.
2. When placing the filter back on again, take care not to damage or bend the wires of the Pt100 sensor
(Figure 7).
Figure 7: Probe head (filter removed)
8.4 Validation of the Analog Outputs
The HM4 offers the possibility to set the analog outputs on a fixed value. This makes it possible to test if the
output stages of the HM4 are working correctly. The function can be activated with the HW4 PC software. Setting
of fixed values is described in the manual E-M-HW4v3-F2-022.