Rotronic HC2 Instrukcja Użytkownika Strona 15

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 23
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 14
E-M-HC2-accessories_24
Rotronic AG
Bassersdorf, Switzerland
Document code Unit
Accessories and Parts
for probes, indicators and transmitters
Instruction Manual
Document Type
Page
15 of 23
Document title
© 2009-2010; Rotronic AG E-M-HC2-accessories_23
Specifications
Supply voltage to AC3011 converter 12…24 VDC (can be used to power the RS-485
slave devices when mA rating is sufficient)
Current consumption (AC3011 only)
85 mA typical
Temperature operating range
-40…85°C
RS-485 specifications
Baud rate : 19200
Parity : none
Data bits : 8
Stop bits : 1
7.2.1 RJ-45 Connector and terminal blocks
The RJ-45 connector is accessible without having to open the AC3011 enclosure.
Terminal Block K1 (Power)
Function
K1-1
GND: Supply voltage (-) to AC3011
K1-2
V + :Supply voltage (+) to AC3011 12 to 24 VDC
K1-3
Earth Ground : Safety ground
Terminal Block K3 (RS-485)
Function
K3-4
D - : TX- / RX- (RS-485 bidirectional)
K3-3
D +: TX+ / RX+ (RS-485 bidirectional)
K3-2
GND: Supply voltage (-) to RS-485 data line (optional)
K3-1
V + : Supply voltage (+) to RS-485 data line 12 to 24 VDC (optional)
Przeglądanie stron 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 21 22 23

Komentarze do niniejszej Instrukcji

Brak uwag